首页 >

高速高焓流动摩阻天平研究与应用

张彭豪,汪运鹏   

  1. 中国科学院力学研究所
  • 收稿日期:2026-06-26 修回日期:2026-10-16 出版日期:2026-10-21 发布日期:2026-10-21
  • 通讯作者: 汪运鹏
  • 基金资助:
    中国科学院基础与交叉前沿科研先导专项

Research and application of skin friction balance for high-speed and high-enthalpy flow

Peng-Hao ZHANG,Yun-Peng WANG   

  • Received:2026-06-26 Revised:2026-10-16 Online:2026-10-21 Published:2026-10-21
  • Contact: Yun-Peng WANG

摘要: 表面摩擦阻力的精确测量对高超声速飞行器气动性能优化至关重要,其测量精度直接影响飞行器气动力特性与发动机推力的准确评估。针对高速高焓流动摩阻测量难题,发展了三种类型的摩阻天平,分别为I型、II-A型和II-B型摩阻天平。提出了采用增材制造技术的一体式摩阻天平加工方案,并应用于II-A型和II-B型摩阻天平。一体式摩阻天平可以与待测表面一体打磨,消除了传统装配式摩阻天平的平齐度误差。因此,传统的摩阻天平测量性能评估中,考虑平齐度因素的不确定度分析方法不再适用。为此提出了改进的不确定度分析方法,系统识别了校准误差、结构误差和信号误差并给出合成不确定度的计算方法。通过数值模拟量化了以往被忽略的唇口力误差,给出了结构误差的保守估计值。为验证上述技术路线的可靠性,利用I型和II-A型摩阻天平在JF-12激波风洞中开展了平板摩阻测量实验,来流的名义马赫数为7,单位雷诺数为10^6 m-1。在95%置信度下,I型和II-A型摩阻天平的校准不确定度均低于0.2%FS(Full Scale,满量程),其中II-A型摩阻天平的测量不确定度为8.9%R(Reading,实际测量的读数值),优于I型摩阻天平的9.4%R,证明了一体式摩阻天平的测量优势。针对超声速燃烧流动环境下壁面摩阻直接测量的技术需求,在原有II-A型摩阻天平基础上,研发了兼具大量程与小型化特性的II-B型摩阻天平,可直接适配超燃冲压发动机内流道的嵌入安装要求。该天平一阶固有振动频率约165 Hz,95%置信度下静态校准不确定度0.81%FS,具备优异的动态响应性能与测量精度。依托该天平在矩形超燃燃烧室试验件中开展了系统的壁面摩阻测量研究,冷态工况下的测量结果与数值模拟、经典经验公式预测结果吻合良好,95%置信度下总测量不确定度稳定处于5.8%R-11.4%R区间,验证了测量系统的可靠性。热态燃烧工况下直接捕捉到边界层燃烧引发的壁面摩阻突变,变化规律与理论预期一致,验证了该型天平在超燃燃烧环境下开展摩阻直接测量的可行性,为超燃流道摩阻特性的精细化研究提供了关键传感器技术支撑。多型适配不同极端工况的专用摩阻天平研制与工程验证,为高速高焓极端流动环境中的壁面剪切力高精度定量测试提供了核心传感技术基础与可信试验数据支撑。

关键词: 一体式摩阻天平, 高速高焓流动, 风洞实验, 壁面剪切力测量, 不确定度分析

Abstract: The precise measurement of skin friction is crucial for the aerodynamic performance optimization of hypersonic vehicles, and the measurement accuracy directly affects the accurate evaluation of the aerodynamic characteristics of aircraft and the engine thrust. In order to solve the problem of skin friction measurements in high-speed and high-enthalpy flows, three types of skin friction balance (SFB) were developed, namely the type I SFB, type II-A SFB and type II-B SFB. A processing scheme for integrated SFB using additive manufacturing technology was proposed and applied to the type II-A and type II-B SFBs. The integrated SFBs can be polished integrally with the test surface, eliminating the misalignment error inherent in traditional assembled SFBs. Therefore, the traditional uncertainty analysis method considering the misalignment in the performance evaluation of SFB measurement is not applicable. An improved uncertainty analysis method was proposed, which systematically identifies the calibration error, structure errors, and signal error, and provides a calculation approach for the combined uncertainty. The previously neglected lip force error was quantified through numerical simulation, and conservative estimates of the structure error were given. To verify the reliability of the above technical approach, the flat plate skin friction measurement experiments were conducted in the JF-12 shock tunnel with the type I and type II-A SFBs. The nominal Mach number of the incoming flow was 7, and the unit Reynolds number was 10^6 m-1. At the 95% confidence level, the calibration uncertainties of the type I and type II-A SFBs are both below 0.2% FS (Full Scale). The measurement uncertainty of the type II-A SFBs is 8.9%R (Reading), outperforming the 9.4%R of the type I SFBs, demonstrating the measurement advantage of the integrated SFBs. In response to the technical demand for direct wall shear stress measurement in supersonic combustion flow environments, the type II-B SFB with both large range and miniaturization characteristics has been developed based on the type II-A SFB, which can be directly adapted to the embedded installation requirements of the internal flow channel of scramjet engine. The first-order natural vibration frequency of this balance is approximately 165 Hz, with a static calibration uncertainty of 0.81%FS at 95% confidence level. It exhibits excellent dynamic response performance and measurement accuracy. Based on this balance, systematic wall shear stress measurement research was conducted in a rectangular scramjet combustor test piece. The measurement results under cold-flow conditions are in good agreement with numerical and classical empirical formula prediction results. The total measurement uncertainty is stably within the range of 5.8%R to 11.4%R at 95% confidence level, verifying the reliability of the measurement system. Under hot-flow conditions, the sudden change in skin friction caused by boundary-layer combustion was directly captured, and the variation pattern is consistent with theoretical expectations. This verifies the feasibility of using the type II-B SFB for direct skin friction measurement in supersonic combustion environments, providing key sensor technology support for the detailed study of skin friction characteristics in supersonic combustion flow channels. The development and engineering verification of specialized SFBs capable of adapting to various extreme flow conditions provide a core sensor technology foundation and credible experimental data support for high-precision measuring of wall shear stress in high-speed, high-enthalpy extreme flow environments.

Key words: integrated skin friction balance, high-speed and high-enthalpy flow, wind tunnel experiment, wall shear stress measurement, uncertainty analysis

中图分类号: