导航

ACTA AERONAUTICAET ASTRONAUTICA SINICA ›› 1988, Vol. 10 ›› Issue (10): 525-529.

• 论文 • Previous Articles     Next Articles

METHOD OF CONTAMINATION ERROR ELIMINATION IN HOT-WIRE ANEMOMETERS MEASUREMENTS

Li Yaping1, Lin Qixun1, Du Qinfang2, Ma Jieping2   

  1. 1. Wuxi Second Mechanical Research Institute;2. Northwestern Polytechnical University
  • Received:1987-07-03 Revised:1900-01-01 Online:1988-10-25 Published:1988-10-25

Abstract: This paper presents a method of calibration with actual flow fields to decrease the influences of oil vapor mixing with dust on the measurement result,and to extend application of hotwire anemometer.Using this method,the measurement error due to contamination can ba controled within a acceptable range, as the source of contamination is near steady. A demostration of two-dimensional flow mesurements is given.