针对主动磁浮微推力测量平台在实物研制前难以综合评估磁力非线性、释放稳定性、执行器动态和推力反演不确定度的问题,以3 kg级主动磁浮微推力测量平台为对象,建立了磁力计算、垂向闭环控制、水平推力反演和不确定度传播的综合评估流程。采用二维轴对称静磁场模型获得气隙、电流与轴向磁力的映射关系,并将非线性磁力模型、电流闭环模型和垂向动力学模型耦合,用于分析释放位置误差、执行器能力、位置噪声和控制延迟对悬浮稳定状态的影响。结果表明,在设计气隙3.50 mm、偏置电流0.30 A条件下,轴向磁力为29.430 N;在电流限幅±0.50 A、电流环带宽100 Hz和0.50 s仿真时长下,满足测量悬浮稳定性要求的初始位置误差范围为?47~149 μm。位置噪声和控制延迟对边界附近工况的影响更显著,当位置噪声标准差为5 μm、附加控制延迟为3 ms时,平衡点和负边界附近满足稳定悬浮的样本比例分别为0.91和0.50。水平推力反演分析表明,若忽略推力建立过程和角向黏性阻尼影响,恒定推力简化模型会低估稳态推力;在50 μN推力下,推力作用时间为1、3、5 s对应的相对标准不确定度分别为2.09%、1.34% 和1.33%。所建立的评估流程可为同类主动磁浮微推力测量平台的磁路参数、释放控制、推力作用时间和反演不确定度评估提供参考。
To address the difficulty in comprehensively evaluating magnetic-force nonlinearity, release stability, actuator dynamics, and thrust-inversion uncertainty prior to hardware implementation of active magnetic-levitation microthrust measurement platforms, a 3 kg-class active magnetic-levitation microthrust measurement platform was considered, and a comprehensive evaluation workflow comprising magnetic-force calculation, vertical closed-loop control, horizontal thrust inversion, and uncertainty propagation was established. A two-dimensional axisymmetric magnetostatic model was used to obtain the mapping relationship among gap, current, and axial magnetic force. The nonlinear magnetic-force model, current closed-loop model, and vertical dynamic model were then coupled to analyze the effects of release-position error, actuator capability, position noise, and control delay on stable levitation. The results show that, at a design gap of 3.50 mm and a bias current of 0.30 A, the axial magnetic force is 29.430 N. With a current limit of ±0.50 A, a current-loop bandwidth of 100 Hz, and a simulation duration of 0.50 s, the initial position-error range satisfying the stable-levitation requirement for measurement is ?47 to 149 μm. Position noise and control delay have a more pronounced effect near the stability boundary. When the position-noise standard deviation is 5 μm and the additional control delay is 3 ms, the proportions of samples satisfying stable levitation are 0.91 at the equilibrium point and 0.50 near the negative boundary. Horizontal thrust-inversion analysis shows that, when thrust buildup and angular viscous damping are neglected, the simplified constant-thrust model underestimates the steady-state thrust. At a thrust of 50 μN, the relative standard uncertainties for thrust application durations of 1, 3, and 5 s are 2.09%, 1.34%, and 1.33%, respectively. The proposed evaluation workflow can provide a reference for magnetic-circuit parameter selection, release control, thrust application duration selection, and thrust-inversion uncertainty evaluation of similar active magnetic-levitation microthrust measurement platforms.